C-Top
Benchtop incubator for up to date,
individual cell cultivation
DESCRIPTION
Advantages of Labo C-Top:
Approved medical device
Two independent and separate incubation chambers;
Parameters (gas flow, temperature) can be adjusted individually
Optimized gassing with premixed gas
Optimized temperature distribution by double heated chamber from top and bottom;
Very fast heating up and recovery times
Reliable humidity system;
Current value of relative humidity is displayed (% rH)
Convenient glass control panel
Space saving
FEATURES
Incubation chambers
Aluminum milled and eloxated
In the lower part:
Slots provide full contact to culture dishes
Adaption for humidity block
Gas inlet
Easy cleaning
Independent, for use with (per chamber):
4 x IVF 4-well dishes 66x66 mm
4 x Petri dishes Ø 60 mm
10 x Petri dishes Ø 35 mm
Temperature
Overall heating from top and bottom
Very homogeneous temperature distribution
Humidity
Humidification by means of humidity block
Reservoir sufficient for 4 days (when gassing with 20 ml/min)
Humidity block autoclavable
Humidity alarm disengageable
Current value of relative humidity is displayed (% rH)
Gassing
Connection of premixed gas (CO2, O2 < 21 %, N2)
Separate measuring and regulation of gas flow for each chamber
Automatic gas purge when closing the chamber
Recovery times
Short recovery times for all parameters through microprocessor controlled regulation
Use of glass panel
Homogeneous glass cover
Sensoric buttons integrated in glass
Image of placements for marking inserted dishes at the glass cover
Easy operation and menu navigation
Lit LCD at working place height
Trouble check system
Optical and acoustic alarm in case of set point deviation or defective sensor
Independent over-temperature protection
Remote alarm
Options
PC software LaboDat+ for documentation of incubator’s parameters
Second gas inlet for use of different gas mixtures for each chamber
Lockable lids
Interfaces
Remote alarm (potential free)
RS485-interface for data logging with optional available PC software LaboDat+
SPECIFICATIONS
Exterior
Dimensions 403 x 225 x 155 mm (w x d x h)
Weight 8.8 kg
Interior
Two chambers with 0.5 liters each
Mixed load possible
Temperature
Overall heating from top and bottom of each chamber
Range 30 °C - 42 °C from 7 °C above
ambient temperature
Stability / uniformity ± 0.1 °C / ± 0.3 °C
Gassing
Premixed gas
Gas flow adjustable between 20 ml/min and 30 ml/min
Flow rate after opening the lid (automatic gas purge):
250 ml/min for 3 minutes
Power Supply
110 – 230 V AC, 50/60 Hz, 100 VA
Primary pressure for gas 0.8 bar
Ambient temperature 18 °C - 30 °C
Classification
Safety class II
Class IIa for every usage according to EC-Directive 93/42/EEC
EN 61010 conform
<
>
C16 Incubator
Incubator with an inner volume of 16 liters
DESCRIPTION
Advantages of C60:
Approved medical device
Active sterile humidity:
External water reservoir
Distilled water is vaporized at 120 °C
Minimized risk of contamination
Humidity sensor (rH)
Very fast recovery
CO2 sensor:
Dual beam infrared sensor
Very precise measurement
Very fast recovery
Long-term stable
Direct heating:
Independent six-sided direct heating
Homogeneous temperature distribution in incubation chamber
Very fast recovery
Optional:
O2-control
Subdivided inner glass door with 4 small doors
Lockable front door
Access port
PC software LaboDat+
SPECIFICATIONS
Exterior Dimensions 510 x 540 x 720 mm (w x d x h)
Interior Chamber volume 60 l
Electro-polished stainless steel
Shelves height adjustable
Temperature Panel heated interior chamber and door
Range 27-42 °C starting 5 °C above ambient temperature
Stability / uniformity ±0.1 °C / ±0.3 °C
Over-temperature alarm 32-47 °C
CO2 control Dual beam infrared sensor
Range 0–10 % CO2
Stability ±0.1 % CO2
O2 control < 21 % (optional) Measurement with galvanic sensor
Connection of N2
Adjustable range 1–21 % O2
Humidity Active sterile humidity generated by vaporizing module at 120 °C
Measuring range 0–98 % rH / range 60–95 % rH
Supply 230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 160 W
Inlet pressure for gas 0.8 bar
Ambient temperature 18-30 °C
Classification
Safety class I
Class IIa for all usage according to EC-Directive 93/42/EE
Conform to ENN 61010
FEATURES
Interior
Electro-polished stainless steel with coved corners
Shelves and ductwork easily removable for cleaning purposes
Temperature Panel heated interior chamber and door
Humidity supply
Internal water reservoir
CO₂
Dual beam infrared sensor
Measurements independent from temperature and humidity
Very short recovery times
For all adjustable parameters through optimized microprocessor control
Easy Handling
Menu guided
Backlit LCD display
Foil keypad
Diagnostic system
Optical and acoustical alarm if set point deviation or sensor defect occurs
Door monitoring
Air monitoring
Hardware self-test
Remote alarm
Options
O₂ control: connection for N₂
Interfaces
Remote alarm (potential free)
RS485-interface for data-logging with optional PC software LaboDat+
<
>
C60 Incubator
Incubator with an inner volume of 60 liters
DESCRIPTION
Advantages of C60:
Approved medical device
Active sterile humidity:
External water reservoir
Distilled water is vaporized at 120 °C
Minimized risk of contamination
Humidity sensor (rH)
Very fast recovery
CO2 sensor:
Dual beam infrared sensor
Very precise measurement
Very fast recovery
Long-term stable
Direct heating:
Independent six-sided direct heating
Homogeneous temperature distribution in incubation chamber
Very fast recovery
Optional:
O2-control
Subdivided inner glass door with 4 small doors
Lockable front door
Access port
PC software LaboDat+
FEATURES
Interior
Electro-polished stainless steel with coved corners
No tools required for quick assembly and disassembly of interior
Temperature
Panel heated chamber and door
Intelligent temperature control system avoids condensation in inner chamber
Active sterile Humidity supply
External water supply container
CO2 control
Dual beam infrared sensor
Measurement independent from temperature and humidity
Fast recovery
For all adjustable parameters through optimized microprocessor control
Easy handling
Menu guided
Large, backlit LCD
Foil keyboard
Diagnostic system Visual and acoustic alarm if set point deviation or sensor defect occurs
Door monitoring
Remote alarm
Options
O2 control: connection of N2
Subdivided inner glass door with 4 small doors
Lockable front door
Access port Ø 30 mm
Interfaces
Alarm contact (potential free)
RS485 interface for connection of optional available PC software LaboDat+
SPECIFICATIONS
Exterior Dimensions 510 x 540 x 720 mm (w x d x h)
Interior Chamber volume 60 l
Electro-polished stainless steel
Shelves height adjustable
Temperature Panel heated interior chamber and door
Range 27-42 °C starting 5 °C above ambient temperature
Stability / uniformity ±0.1 °C / ±0.3 °C
Over-temperature alarm 32-47 °C
CO2 control Dual beam infrared sensor
Range 0–10 % CO2
Stability ±0.1 % CO2
O2 control < 21 % (optional) Measurement with galvanic sensor
Connection of N2
Adjustable range 1–21 % O2
Humidity Active sterile humidity generated by vaporizing module at 120 °C
Measuring range 0–98 % rH / range 60–95 % rH
Supply 230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 160 W
Inlet pressure for gas 0.8 bar
Ambient temperature 18-30 °C
Classification
Safety class I
Class IIa for all usage according to EC-Directive 93/42/EE
Conform to ENN 61010
<
>
C200
Incubator with an inner volume of 200 liters
DESCRIPTION
Advantages of C200:
Approved medical device
Active sterile humidity:
External water reservoir
Disteilled water is vaporized at 120 °C
Minimized risk of contamination
Humidity sensor (rH)
Very fast recovery
CO2 sensor:
Dual beam infrared sensor
Very precise measurement
Very fast recovery
Long-term stable
Direct heating:
Independent six-sided direct heating
Homogeneous temperature distribution in incubation chamber
Very fast recovery
Optional:
O2 control
Subdivided inner glass door with 6 small doors
Lockable front door
Access port
PC software LaboDat+
FEATURES
Interior
Electro-polished stainless steel with coved corners
No tools required for quick disassembly and assembly of interior components
Temperature
Panel heated chamber and door
Intelligent temperature control system avoids condensation in inner chamber
Active sterile Humidity supply
External water supply container
CO2 control
Dual beam infrared sensor
Measurement independent from temperature and humidity
Fast recovery
For all adjustable parameters through optimized microprocessor control
Easy handling
Menu guided
Large, backlit LCD
Foil keyboard
Diagnostic system
Optical and acoustic alarm if set point deviation or sensor defect occurs
Door monitoring
Remote alarm
Options
O2 control: connection for N2 possible
Subdivided inner glass door with 6 small doors
Lockable front door
Access port Ø 30 mm
Interfaces
Remote alarm (potential free)
RS485 interface for connection of optional available PC software LaboDat+
SPECIFICATIONS
Exterior Dimensions 680 x 690 x 930 mm (w x d x h)
Interior Chamber volume 200 l
Electro-polished stainless steel
Shelves height adjustable
Temperature Panel heated interior chamber and door
Range 27-42 °C starting 5 °C above ambient temperature
Stability / uniformity ±0.1 °C / 0.3 °C
Over-temperature alarm 32-47 °C
CO2 control Dual beam infrared sensor
Range 0–10 % CO2
Stability ±0.1 % CO2
O2 control < 21 % (optional) Measurement with galvanic sensor
Connection of N2
Adjustable range 1–21 % O2
Humidity Active sterile humidity generated by vaporizing module operating at 120 °C
Measuring range 0–98 % rH / adjustable range 60–95 % rH
Supply 230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 200 W
Inlet pressure 0.8 bar
Ambient temperature 18-30 °C
Classification Safety class I
Class IIa for all usage according to EC-Directive 93/42/EEC
Conform to EN 61010
<
>
Cell Trans 2018
Transport incubator
DESCRIPTION
Transport incubator for a secure and mobile preservation of specimen, culture media etc.
FEATURES
Secure transport of cells without electric mains and CO2 connection
Factory-aligned parameters: 37 °C and 5 % CO2
Different temperature or CO2 concentration possible (to be advised before order)
Operation at 12 V car voltage possible
Use of different heating blocks
Easy handling and cleaning
Compact Design
Parameters* during transport without connection to electric mains and CO2 bottle:
Time Temp. CO2
0 min 37 °C 5 %
10 min 36.6 °C
4.9 %
30 min 35.5 °C 4.8 %
60 min 34.0 °C 4.7 %
*The measurements relate to an ambient temperature of 23 °C .
SPECIFICATIONS
Power supply
115 V or 240 V AC, 50/60 Hz
12 V DC with car voltage via adapter cable
Dimensions
Outside: 385 x 245 x 305 mm (l x w x h)
Inside: approx. 312 x 167 x 140 mm (l x w x h)
Weight
Approx. 4.7 kg
Temperature
Available between 32 °C and 39 °C (to be advised before order)
Accuracy of regulation
± 0.3 °C (measured with heating block REF 10264)
Heating up time
Approx. 2 h
CO2 concentration
5 % (different concentrations possible)
Heating block inside necessary
Connection to CO2 supply with pre-pressure of 0.8 bar necessary
CO2 accuracy ± 0.3 % CO2 (when supplied with CO2 with pre-pressure of 0.8 bar ± 0.05 bar)
<
>
Cell Trans +
CO2 transport incubator for a secure and mobile preservation of specimen, culture media etc.
DESCRIPTION
Advantages of CellTrans+:
Can also be used as back up in case of incubator or power outage
Temperature adjustable between 30 °C and 42 °C
CO2 adjustable between 0 % and 10 %
Reliable humidity system
Easy handling and cleaning
Compact design
Insert for 16 Falcon tubes of 5ml and 13ml each
Duration with completely charged accumulator:
At 2 °C ambient temperature approx. 4 h
At 10 °C ambient temperature approx. 6 h
At 23 °C ambient temperature approx. 10 h
FEATURES
Interior chamber
Anodized aluminum
Insert: stainless steel
Temperature
Over-all heating of lateral surface, bottom and lid of incubation chamber
Homogeneous temperature distribution
Humidity
Via humidifying disc Ø 50 mm
CO2
Dual beam infrared sensor
Measurements independent of temperature and humidity
Recovery times
Short recovery times for all adjustable parameters through microprocessor controlled regulation
Easy navigation
Menue guided
Backlid LCD
Interfaces
USB interace for data evaluation via PC Software DataVISUAL ’19 (in scope of supply) (Compatible up to Windows 10)
SPECIFICATIONS
Power supply
100 V – 240 V, 60 W
12 V car cable
Dimensions Exterior:
386 x 251 x 370 mm (l x w x h in mm)
Interior:
ø 160 mm, Height 130 mm, volume approx. 1.7 l
Empty weight approx. 10 kgs.
Temperature control
Control range 30 °C – 42 °C at 3 °C above room temperature
Temperature homogeneity ± 0.2 °C
Temperature stability ± 0.1 °C
CO2 regulation
Control range 0 – 10 % CO2
Control accuracy 0.3 % CO2
CO2 pressure tank
Capacity 0.4 l
Pressure 8 bar
Classification
Safety class I
Conform to EN 61010
<
>
Thermo Cell
Incubator Controller with Available Case
DESCRIPTION
Transport Incubator for Safe & Versatile Use
FEATURES
•Factory-aligned temperature at 37 °C
•Different temperature possible (to be advised before order)
•Continuous operation with power supply, car voltage or integrated accumulator during transport
•Operation with batteries up to 5 hours (at 22 °C ambient temperature)
•For use with 8 x 13 ml Falcon tubes (REF 13881) or 2 x 13 ml and 8 x 5 ml Falcon tubes (REF 16518)
•Compact design
SPECIFICATIONS
Dimensions 111 x 111 x 215 mm (w x d x h)
Weight 2.4 kg
Power supply 100 - 240 V AC, 50/60 Hz
Temperature Between 32 °C and 42 °C available
(to be advised before order)
Accuracy of regulation ± 0.3 °C
Heating up time 22 °C to 37 °C approx. 1 h*
Operation time With power supply or 12 V car voltage continuously
With accumulator approx. 5 h*
Charging time of accumulator Approx. 3 h*
Heating block For use with 8 x 13 ml Falcon tubes (REF 13881) or 2 x 13 ml and 8 x 5 ml Falcon tubes (REF 16518)
*at 22 °C ambient temperature
<
>
InControl
Measuring Device, InControl 1050, InControl
DESCRIPTION
Independent measuring device for control and evaluation of CO2, temperature and O2
FEATURES
Accurate measuring of CO2, temperature and optional O2
Single or continuous measurement in arbitrary intervals
Measurement and documentation of multiple incubators
Documentation of measurements with date, time and incubator number
Easy menu navigation
Data-Download via USB; PC software DataVISUAL`09 incl.
Continuous operation by Li-Ion battery or mains adapter possible
SPECIFICATIONS
Case Plastic (ABS)
CO2 measurement Dual beam infrared sensor
Range: 0 - 10 % CO2
Display definition: 0.1 % CO2
Accuracy:
0 - 6 % CO2 ± 0.2 % CO2
6 % - 10 % CO2 ± 0.3 % CO2
Sample volume per measurement (2 min.): approx. 0.8 l
Temperature measurement Platinum temperature sensor PT1000
Range: 0 - 100 °C
Display definition: 0.1 °C
Accuracy:
20 °C - 50 °C ± 0.2 °C
50 °C - 100 °C ± 0.3 °C
O2 measurement Galvanic sensor
Range: 0 - 100 % O2
Display definition: 0.1 % O2
O2 sensor not included
Data logger
Recording of measurement logging in arbitrary intervals
Output of measurements via:
backlit LCD
PC software Data Visual’09 (via USB data cable)
Interval for data logging: 15 - 120 min., adjustable in steps of 5 min.
Storage of max. 1008 measurements with date and time
Power supply Rechargeable Li-Ion battery
Power supply 100 – 240 V AC, 5 V DC 1.5 A
Operation time Use of rechargeable battery approx. 4 h
Time to recharge battery approx. 4 h
<
>
Hot Plates A3, A4
Large hotplates for laboratory and scientific use
DESCRIPTION A3 HOT PLATE
Advantages:
Temperature control between 27 °C and 45 °C
Homogeneous temperature distribution
Planar surface for optimal heat transfer to heating blocks, culture dishes and slides
Easy to clean glass surface
Short heating up time
Extremely flat with bigger work surface
Space saving due to separate control unit
SPECIFICATIONS A3 HOT PLATE
Dimensions 420 x 300 x 23 mm (w x d x h)
Weight 5.5 kg
Temperature Adjustable between 27 °C and 45 °C
Heating up time < 15 min.
Power supply 100 – 240 V AC, 50/60 Hz, 140 W
Classification Protection class IPX1 (Hot Plate)
Protection class IPX0 (Controller)
Safety class I
Cable length
Between Controller and Hot Plate 1.5 m
DESCRIPTION A4 HOT PLATE
Advantages:Robust and planar glass surfaceHomogeneous temperature distributionOLED display for intuitive and easy-to-use user menuFunctional buttons with haptic feedback via vibration (disengageable)Extremely flat and space savingCleaning and disinfection with all common detergents
SPECIFICATIONS A4 HOT PLATE
Dimensions 300 x 210 x 15 mm (w x d x h)
Usable surface area 255 x 140 mm
Weight 1050 g (w/o mains adapter)
Temperature stability ±0.2 °C
Temperature homogeneity ±0.3 °C
Temperature Adjustable between 27-45 °C (2 °C above ambient temperature)
Heating up time Approx. 10 min.
Power supply 100 – 240 V AC, 60 W
Cable length Hotplate to mains adapter 1.8 m, mains adapter to plug 1.8 m
NextGen LifeLabs provides specialized, cutting-edge equipment, services, and support for the modern A.R.T. laboratory. The ONLY Company that customizes equipment to meet your needs.
CONTACT US
384 Nina Way
Warminster, PA 18974
Office: 215-672-4838
Email: info@nextgenlifelabs.com
Fax: 215-672-4904
CONTACT US